|
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number, d9: device availability, h2: if follow-up, what type, h3: device evaluated by manufacturer, h4: device manufacturer date, h6: type of investigation, h6: investigation findings, h6: investigation conclusions, h10: additional narratives/data.Zimvie received one implant for evaluation.Visual evaluation / functional test was performed, implant shows slight wear however mount was able to disengage from implant as intended.No malfunction.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was torque or speed applied during placement/seating exceeds recommended value.Therefore, based on the available information, a device malfunction did not occur.The mount was able to disengage from implant as intended.The reported is unconfirmed however the coob could not be verified since the condition of the product/packaging when received by the customer was unknown.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|