| Model Number |
863279 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported that the equipment came off the cart and fell to the floor.The device was not in use on a patient at the time of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips remote service engineer (rse) remotely interviewed the customer who was onsite.The customer confirmed the torx screw was loose.The rse dispatched a philips field service engineer (fse) to the location.The fse checked the equipment, and the anchoring system was adjusted to the rolling cart to resolve the issue.The fse noted there was nothing abnormal, apart from a missing screw and loose screws on cart fixation to suresign.Based on the information available and the testing conducted, the cause of the reported problem was confirmed to be the anchoring system.After the fse evaluated the unit and the anchoring system was adjusted to the rolling cart, the device was returned to functional use with no further issues identified.The device remains at the customer site.No further investigation or action is warranted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Data correction to device identifier.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 product udi: the manufacturing date for the reported device is prior to 24sept2016 so no udi required.
|
| |
|