| Catalog Number |
1804400-23 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the mid left anterior descending (mlad) artery with heavy calcification and moderate tortuosity.The 4.0x23mm xience skypoint stent delivery system (sds) had resistance during advancement and failed to cross due to the anatomy.There was resistance with the anatomy during removal but was removed independently.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|