| Model Number |
UNICEL DXH SLIDEMAKER STAINER |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The bec employee (field service engineer, fse) reported that while troubleshooting on dxh sms, he suffered a lower back injury when reinstalling the dxh sms sample transport module (stm).The fse stated he immediately notified his manager and called the employee nurse hotline.The hotline scheduled a doctor's appointment for next day on ((b)(6) 2023).The fse stated he was administered a toradol injection and prescribed methyl prednisolone, and the doctor requested an mri to evaluate the severity of the condition.The fse was out for two weeks on workman's compensation.
|
| |
|
Additional Manufacturer Narrative
|
|
In a follow up communication, the fse stated his condition has been improving and there has been no complications since initial incident.The injury was not a result of a malfunction.The event occurred as a result of user error, the fse should not have attempted to lift the heavy part by himself.Per service manual pn 4237476 current rev, section 3.19 [installation on bench top] moving spm, unpacking shipping crate, and checking accessory kit.Risk of operator injury, lower back, or pinch injury.The weight of the spm with the stm module removed is 77 kg (170 lbs.).Exercise caution and follow all published ehs guidelines.To prevent injury, the lifting handles must be installed, and a minimum of two people are required to safely lift the analyzer.Bec internal identifier: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up 01: -section d4 catalog number - corrected to reflect instrument number.Original report reflected component part number from the instrument.-section h4 device manufacture date - updated to reflect correct manufacturing date.
|
| |
|
Search Alerts/Recalls
|