| Catalog Number |
1009666J |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Difficult to Remove (1528); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the circumflex (cx) artery with mild calcification and mild tortuosity.The hi-torque (ht) balance middleweight (bmw) universal ii guide wire was noted to have delaminated catching on the ultrasound guided catheter.The black polymer coating was noted to have peeled.The guide wire and the ultrasound guided catheter were removed as a unit and managing to be removed with abrupt force.Another non-abbott guide wire was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d9 - device available for evaluation updated from ¿yes¿ to ¿no¿.The device was returned for analysis.Electronic lot history record (elhr) and exception reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot level product quality issue.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the ultrasound guided catheter was due to the mildly calcified and mildly tortuous anatomy and/or a coagulation of blood/contrast on the guide wire resulted in the reported difficult to advance and the reported difficult to remove.Interaction and/or manipulation of the device(s) resulted in the reported peeled black polymer coating.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported peeled / delaminated was unable to be confirmed.The reported difficult to advance and the reported difficult to remove were unable to be replicated in a testing environment as they were based on operational circumstances.Electronic lot history record (elhr) and exception reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot level product quality issue.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the ultrasound guided catheter due to the mildly calcified and mildly tortuous anatomy and/or a coagulation of blood/contrast on the guide wire resulted in the reported difficult to advance and the reported difficult to remove.Inadvertent mishandling and/or interaction with the mildly calcified and mildly tortuous anatomy resulted in the noted multiple core bends likely contributing to the reported difficulties.The reported peeled / delaminated was unable to be confirmed.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D9 - device available for evaluation updated from no to yes.Correction to h6: component code 4755 removed; 4721 added.Type of investigation code 4114 removed; 10 added.
|
| |
|