| Model Number |
GA-2003477 |
| Medical Device Problem Codes |
Defective Device (2588); Output below Specifications (3004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the blast shield lens was burnt during the procedure and the blast shield lens and probe were replaced but the output seems weak.The procedure was completed with a non-bsc product without patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
The console was not returned for analysis; however, a field service engineer (fse), conducted an onsite visual inspection of the console.While performing the visual inspection the fse found the probe and two blast shields to be defective.After replacing the two blast shield lens assemblies, performing a laser optical axis adjustment, and output check the reported complaint of low power output was resolved.Based on the field service performed, it is most likely that the damaged probe and debris shields could have affected the device performance thus confirming the reported finding.H3 other text : device remains at health care facility.
|
| |
|
Event or Problem Description
|
|
It was reported that the blast shield lens was burnt during the procedure and the blast shield lens and probe were replaced but the output seems weak.The procedure was completed with a non-bsc product without patient complications.
|
| |
|