• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD MOSES PULSE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD MOSES PULSE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number GA-2003477
Medical Device Problem Codes Defective Device (2588); Output below Specifications (3004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the blast shield lens was burnt during the procedure and the blast shield lens and probe were replaced but the output seems weak.The procedure was completed with a non-bsc product without patient complications.
 
Additional Manufacturer Narrative
The console was not returned for analysis; however, a field service engineer (fse), conducted an onsite visual inspection of the console.While performing the visual inspection the fse found the probe and two blast shields to be defective.After replacing the two blast shield lens assemblies, performing a laser optical axis adjustment, and output check the reported complaint of low power output was resolved.Based on the field service performed, it is most likely that the damaged probe and debris shields could have affected the device performance thus confirming the reported finding.H3 other text : device remains at health care facility.
 
Event or Problem Description
It was reported that the blast shield lens was burnt during the procedure and the blast shield lens and probe were replaced but the output seems weak.The procedure was completed with a non-bsc product without patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSES PULSE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidman st. 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
yokneam industrial park
hakidm
yokneam 20692 04
IS   2069204
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key17909371
Report Number2124215-2023-55968
Device Sequence Number6653284
Product Code GEX
UDI-Device Identifier07290109145440
UDI-Public07290109145440
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGA-2003477
Device Catalogue NumberGA-2003477
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer10/30/2023
Initial Report FDA Received Date10/10/2023
Supplement Report FDA Received Date11/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/16/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-