• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH CORTEX SCREW AXSOS 3 Ø4.5MM / L60MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH CORTEX SCREW AXSOS 3 Ø4.5MM / L60MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 661760
Medical Device Problem Codes Material Deformation (2976); Material Twisted/Bent (2981)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.H3 other text: device remains implanted in the patient.
 
Event or Problem Description
Surgeon implanted a 355mm distal plate using the mis technique during an eps case.His feedback as reported: "4.5 non locking cortex not strong - bent upon insertion." "stronger cortical screws; distal rim on the plates too wide." "see above.Bulky outrigger does not lock to plate.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.¿ this eps feedback was forwarded to rnd for review with following result: "4.5 non locking cortex not strong - bent upon insertion." [provided as a comment on the variety of offered screw types and sizes] "stronger cortical screws; distal rim on the plates too wide." [indicated as identified area for improvement] "see above.Bulky outrigger does not lock to plate." [indicated as a drawback] "when inserted, the cortex screws have a tactile feeling that is softer than the locking screws because of the different material (ti6al4v vs.Cr-co).However, this difference ultimately has no influence on the effectiveness of the cortex screws that have already been successfully used for axsos plate system.The plate thickness is driven by the locking hole height, the plate width was chosen to accommodate for as many screw hole options as possible while avoiding a potential total knee component.No updates/modifications to plate thickness/shape planned as of now." more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
Surgeon implanted a 355mm distal plate using the mis technique during an eps case.His feedback as reported: "4.5 non locking cortex not strong - bent upon insertion." "stronger cortical screws; distal rim on the plates too wide." "see above.Bulky outrigger does not lock to plate.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTEX SCREW AXSOS 3 Ø4.5MM / L60MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17909489
Report Number0008031020-2023-00355
Device Sequence Number18023893
Product Code HRS
UDI-Device Identifier07613252580768
UDI-Public07613252580768
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K222381
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number661760
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/14/2023
Supplement Date Received by Manufacturer10/31/2023
Initial Report FDA Received Date10/10/2023
Supplement Report FDA Received Date11/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-