MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
VNMC3428C207TU |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A valiant navion stent graft was implanted in the treatment on a taa on the (b)(6) 2020.4 valiant navion stent grafts implanted in total, from proximal to distal vnmc3434c90tu, vnmc3131c223tu,vnmc2525c94tu were implanted on the (b)(6) 2020.The final navion vnmc3428c207tu was implanted one month later for the abdominal aorta for the treatment of a aaa.It was reported that a follow-up ct taken on the (b)(6) 2023 showed a type iiib endoleak in the navion stent graftvnmc3131c223tu v29818675.Core lab review of ct imaging from (b)(6) 2023 identified a new type iiib endoleak in vnmc3131c223tu (v29818675) and vnmc3428c207tu (v29961271).A new stent fracture was identified on ring 4 of vnmc3131c223tu (v29818675).New sre was reported of +4.3mm on ring 4 of vnmc3131c223tu (v29818675) and + 2.0mm to ring 5 of vnmc3131c223tu (v29818675).No stent ring migration was observed.The max aortic diameter was noted as 70.9mm.The imaging was noted as na for aortic enlargement and non evaluable for some grafts for stent fractures and stent ring migration due to device overlaps.The stent graft was relined using a valiant captivia stent graft and the endoleak was confirned as resolved.Per the physician the cause of the event was stent graft failure.No additional clinical sequalae were provided and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued from h9: 3005364322-02-19-2021-001-r.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|