| Catalog Number |
5733730 |
| Medical Device Problem Codes |
Moisture Damage (1405); Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to a dialysis catheter placement procedure, the packaging was allegedly damaged.It was further reported that the inner seal was visibly opened but the contents were intact.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 11/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a dialysis catheter placement procedure, the packaging was allegedly damaged.It was further reported that the inner seal was visibly opened but the contents were intact.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed hemosplit d/l catheter kit was returned for evaluation.Gross visual evaluations were performed.The sample appeared to have minor humidity stains throughout the package.Components within kit did not appear affected.Therefore, the investigation is confirmed for the identified moisture damage problem.However, the investigation is unconfirmed for the reported unsealed device packaging as the tray inside the pouch will be unsealed which refers to the gaps between the top and bottom of the tray.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a dialysis catheter placement procedure, the packaging was allegedly damaged.It was further reported that the inner seal was visibly opened but the contents were intact.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed hemosplit d/l catheter kit was returned for evaluation.Gross visual evaluations were performed.Components within kit did not appear affected.Therefore, the investigation is unconfirmed for the reported unsealed device packaging as the product was returned with the seal intact.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2024), g3 h11: h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|