• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC INSTINCT ENDOSCOPIC HEMOCLIP; PKL, LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC INSTINCT ENDOSCOPIC HEMOCLIP; PKL, LIGATOR, HEMORRHOIDAL Back to Search Results
Catalog Number INSC-7-230-S
Medical Device Problem Codes Material Separation (1562); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
During a polyp clipping, the physician used a cook instinct endoscopic hemoclip.It was reported that the drive wire advanced but retracted when attempting to open clip [moved in a tromboning motion].This device was removed and discarded.Another clip was used to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The system preparation section of the instructions for use includes the following: "uncoil device.Verify smooth handle operation and clip action.Open clip by gently moving handle spool distally (away from handle thumb ring).Once clip is fully open, do not continue advancing handle spool as clip may prematurely detach from catheter.Close clip by moving handle spool proximally until clip is fully closed.Caution: do not continue to pull handle spool beyond tactile resistance as this may prematurely deploy clip." the instructions for use states: "endoscope must remain as straight as possible when inserting or withdrawing device." the instructions for use states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope or colonoscope.Caution: holding handle spool during advancement may prematurely deploy clip." failure to follow the instructions above can result in damage to the device which may lead to premature clip deployment.However, even if the clip is not deployed, damage can occur to internal device components such as the driver legs.Once this damage occurs, the clip is in a partially deployed state and may not be able to be opened/reopened.Prior to distribution, all instinct endoscopic hemoclips are subjected to a visual and functional inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT ENDOSCOPIC HEMOCLIP
Common Device Name
PKL, LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key17912306
Report Number1037905-2023-00494
Device Sequence Number15109979
Product Code PKL
UDI-Device Identifier00827002183438
UDI-Public(01)00827002183438(17)241025(10)W4529624
Combination Product (Y/N)N
PMA/510(K) Number
K132809
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2024
Device Catalogue NumberINSC-7-230-S
Device Lot NumberW4529624
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/18/2023
Initial Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCOPE, UNKNOWN MAKE AND MODEL.
Patient SexUnknown
-
-