EXACTECH, INC. CC TIBIAL INSERT SZ 1, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
208-21-09 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Laxity (4526)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this 75 y/o female patient had a partial revision 6 years post op initial tka.The patient was suffering, was presenting laxity & instability.Synovectomy & lavage were performed.The pe retrieved appeared to be badly damaged for both tibial insert & patella.Surgeon replaced both insert & patella, switching from 9mm to 11mm.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) pending evaluation.(d10) concomitant device(s): (b)(6), 200-02-29 - rotule 3 plots diam 29 6.1mm optetrak, (b)(6), 204-32-08 - tige extension diam 12 lg 80, (b)(6), 204-34-08 - tige extension diam 14 lg 80, (b)(6), 204-62-05 - 1/2 cale tibiale ep 5 t.2, (b)(6), 208-01-01 - femur optetrak cimente cc t.1, (b)(6), 208-04-23 - embase tibiale cimentee decalee alpha, (b)(6), 208-04-70 - vis p/platine decalee aa, (b)(6), 208-06-01 - cale femorale distale ep 10 t.1, (b)(6), 208-09-02 - bague femorale 2 degres.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of instability, which led to prosthesis wear.
|
| |
|
Additional Manufacturer Narrative
|
|
*the following sections have corrected information: (h6) health effect - clinical code, medical device problem.
|
| |
|
Additional Manufacturer Narrative
|
|
*the following sections have corrected information: (h6) type of investigation, investigation findings, investigation conclusion.
|
| |
|