• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CC TIBIAL INSERT SZ 1, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CC TIBIAL INSERT SZ 1, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 208-21-09
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Laxity (4526)
Date of Event 09/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this 75 y/o female patient had a partial revision 6 years post op initial tka.The patient was suffering, was presenting laxity & instability.Synovectomy & lavage were performed.The pe retrieved appeared to be badly damaged for both tibial insert & patella.Surgeon replaced both insert & patella, switching from 9mm to 11mm.No further information provided.
 
Additional Manufacturer Narrative
Section h10: (h3) pending evaluation.(d10) concomitant device(s): (b)(6), 200-02-29 - rotule 3 plots diam 29 6.1mm optetrak, (b)(6), 204-32-08 - tige extension diam 12 lg 80, (b)(6), 204-34-08 - tige extension diam 14 lg 80, (b)(6), 204-62-05 - 1/2 cale tibiale ep 5 t.2, (b)(6), 208-01-01 - femur optetrak cimente cc t.1, (b)(6), 208-04-23 - embase tibiale cimentee decalee alpha, (b)(6), 208-04-70 - vis p/platine decalee aa, (b)(6), 208-06-01 - cale femorale distale ep 10 t.1, (b)(6), 208-09-02 - bague femorale 2 degres.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of instability, which led to prosthesis wear.
 
Additional Manufacturer Narrative
*the following sections have corrected information: (h6) health effect - clinical code, medical device problem.
 
Additional Manufacturer Narrative
*the following sections have corrected information: (h6) type of investigation, investigation findings, investigation conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CC TIBIAL INSERT SZ 1, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key17912967
Report Number1038671-2023-02502
Device Sequence Number7526035
Product Code JWH
UDI-Device Identifier10885862054081
UDI-Public10885862054081
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K954208
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2020
Device Catalogue Number208-21-09
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2023
Supplement Date Received by Manufacturer02/19/2024
08/22/2024
10/18/2024
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date03/04/2024
09/16/2024
11/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age75 YR
Patient SexFemale
-
-