Catalog Number 60-6085-202A
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of a customer that the 60-6085-202a, vcare 200a ¿ large was to be used in a hysterectomy procedure on (b)(6) 2023 and, during pre-operative testing, they found the ¿vcare balloon not deflating¿.There was no report of impact/injury, medical intervention or prolonged hospitalization to the patient.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device has not been returned to date and no photographic evidence has been provided; therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.(b)(4).Per the instructions for use, the user is advised: remove vcare from its sterile packaging and inspect for damage caused by shipping.Discard if any damage is noted.Draw 7-10 cc of air into a standard syringe and insert it into the luer connection at the end of the pilot balloon.Test the intrauterine balloon by injecting air with the syringe and checking to see if balloon remains inflated.If balloon does not remain inflated, do not use.Discard and select another vcare unit.After the successful balloon test, deflate the balloon by removing all air with the syringe.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of a customer that the 60-6085-202a, vcare 200a ¿ large was to be used in a hysterectomy procedure on (b)(6) 2023 and, during pre-operative testing, they found the ¿vcare balloon not deflating¿.There was no report of impact/injury, medical intervention or prolonged hospitalization to the patient.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCARE 200A - LARGE
Common Device Name
CANNULA, MANIPULATOR/INJECTOR, UTERINE
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
beth madeline
11311 concept blvd
largo, FL 33773
7274708358
MDR Report Key17913371
Report Number1320894-2023-00207
Device Sequence Number6777423
Product Code LKF
UDI-Device Identifier10653405003809
UDI-Public(01)10653405003809(17)240828(10)202208291
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K142716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2024
Device Catalogue Number60-6085-202A
Device Lot Number202208291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Supplement Date Received by Manufacturer12/05/2023
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date12/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexFemale
Patient EthnicityNon Hispanic
-
-