Model Number 413570
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "the product broke even before using, (several times), and after using it broke again and was left behind in the wound.They were unable to remove it." per additional information received, the dressing was changed twice a day.Photos depicting the reported complaint issue were provided by the complainant.
 
Additional Manufacturer Narrative
Device 1 of 2.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 1000317571.
 
Additional Manufacturer Narrative
Aquacel ag+ wsf 1x45cm (1x5pk)ster eur was manufactured under system application product (sap) code 1708336 and manufacturing lot number 3f00342 on 02 june 2023.Lot # 3f00342 was sterilized under reference id 2173-38396a and released on review of results of sterilization provided by sterilization company steris.All of the results were within specification and products were released.No nonconformity was identified during the manufacturing process of lot 3f00342.This is the only complaint for the affected lot registered within database.Two photographs were received for this issue so could be evaluated in accordance with work instruction (wi).The photos shows the expected product and complaint issue, where a broken dressing is shown following removal from a wound.The lot number of the product is not shown in the images.Unused complaint samples were requested on 03 october 2023, and were received 16th october 2023.Upon receipt of the complaint samples, several dressings from the packs received were checked for tensile strength.A dry tensile test was completed on the samples, to confirm all dressings met the specification.The tensile testing was completed against specification and test method, with the specification 6n/cm being exceeded with all tested dressings.This confirms all dressings tested have met specification.As no inconsistencies or discrepancies were identified within the batch record, no nonconformance is identified.The dressings have been found to meet specification.The image provided with the complaint has raised some questions with the clinical team.The complainant has been asked the dressing change frequency, as the dressing does not look fully gelled and therefore may be more difficult to remove.The dressing change frequency was identified as twice per day, which was identified as frequent.European customer support were asked why the dressing changes were so frequent, and it was identified from the healthcare professional that this was due to exudate levels, although it would not be ecs advice, but that they do what is considered right at the time.If high exudate levels were not maintained throughout the twice daily dressing changes, it is likely that the dressing has not gelled properly and has become stuck and separated in the wound.The dressing has a tensile specification of 6n/cm, so this force could easily be exceeded when pulling the dressing from the wound if it remains fairly dry.It was confirmed that the dressings had not been cut prior to use, but it is possible for forces above the specification to have been exerted on the dressing when removing it.As no nonconformance was identified, the dressing is thought to be within specification and no further investigation is required.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number, reporting site: 1049092, manufacturing site: 1000317571.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUACEL AG+ RIBBON WSF
Common Device Name
DRESSING, WOUND, DRUG
Manufacturer (Section D)
CONVATEC LTD
first avenue
deeside industrial park
deeside, flintshire CH5 2 NU
UK  CH5 2NU
Manufacturer (Section G)
CONVATEC LTD
first avenue
deeside industrial park
deeside, flintshire CH5 2 NU
UK   CH5 2NU
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key17913696
Report Number1000317571-2023-00288
Device Sequence Number12613758
Product Code FRO
UDI-Device Identifier00768455132058
UDI-Public00768455132058
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2025
Device Model Number413570
Device Lot Number3F00342
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer10/20/2023
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-