• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW MULTILUMEN/PSI KIT 9 FR X 11.5 CM; CATHETER,INTRAVASCULAR,THERAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW MULTILUMEN/PSI KIT 9 FR X 11.5 CM; CATHETER,INTRAVASCULAR,THERAP Back to Search Results
Catalog Number ASK-21242-MMC1
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Extravasation (1842); Scar Tissue (2060)
Date of Event 07/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Medwatch form received reports: "rij 9 fr dl introducer placed via us guidance and removed 2.5 weeks later.Extravasation noted at right ij central venous access catheter site while performing site care; actual date of extravasation not clear as there are multiple episodes of leaking catheter/site with dressing changes performed for several days prior; patient received vasopressors and antibiotics at various intervals.Consult of site revealed no underlying necrotizing soft tissue injury but will need debridement at some time in future.Non obstructive clot of rij seen on ct scan of neck 2 days post removal.Site being monitored.No line saved.".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "clinicians must be aware of complications/undesirable sideeffects associated with this device inclu ding, but not limited to.Extravasation.Warning: to reduce the risk of possible vessel wall perforation, do not leave tissue dilator in place as an indwelling catheter." a device history record review was performed based on a potential lot from sales history and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Corrected data: event date changed to 07/14/2023 after receipt of additional information.Codes added for scar and another device needed.Additional information informed us that the patient was deceased.Time interval between the complaint issue and death was 12 days.Cause of death was respiratory failure.It was stated that the death was not caused by the incident.
 
Event or Problem Description
Medwatch form received reports: "rij 9 fr dl introducer placed via us guidance and removed 2.5 weeks later.Extravasation noted at right ij central venous access catheter site while performing site care; actual date of extravasation not clear as there are multiple episodes of leaking catheter/site with dressing changes performed for several days prior; patient received vasopressors and antibiotics at various intervals.Consult of site revealed no underlying necrotizing soft tissue injury but will need debridement at some time in future.Non obstructive clot of rij seen on ct scan of neck 2 days post removal.Site being monitored.No line saved.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW MULTILUMEN/PSI KIT 9 FR X 11.5 CM
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key17914110
Report Number9680794-2023-00783
Device Sequence Number20496499
Product Code FOZ
UDI-Device Identifier10801902112054
UDI-Public10801902112054
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K011761
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue NumberASK-21242-MMC1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2023
Supplement Date Received by Manufacturer11/02/2023
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VASOPRESSORS AND ANTIBIOTICS; VASOPRESSORS AND ANTIBIOTICS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-