DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX54OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121932054
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
This pinnacle insert was removed from patient after dislocating.During the removal it was discovered that most of the ard tabs were gone.There was no surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : this pinnacle insert is removed from a patient after dislocating.During the removal it was discovered that most of the ard tabs were gone.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that pinn mar neut 32idx54od had 4 (four) of the 6 (six) anti-rotation device (ard) tabs sheared off, fragments were not returned for evaluation.Additionally, deformation was also found on the edge of the liner.Based on the observations, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.A manufacturing record evaluation was performed for the finished device [121932054 / c82986] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified during manufacturing.The overall complaint was confirmed as the observed condition of the pinn mar neut 32idx54od would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device [121932054 / c82986] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified during manufacturing.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX54OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17914214
Report Number1818910-2023-20661
Device Sequence Number7525836
Product Code LPH
UDI-Device Identifier10603295013914
UDI-Public10603295013914
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Catalogue Number121932054
Device Lot NumberC82986
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/28/2023
Supplement Date Received by Manufacturer10/23/2023
11/10/2023
11/14/2023
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date10/24/2023
11/13/2023
11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 32 +5 BR; UNKNOWN HIP ACETABULAR CUP
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-