| Model Number |
WA22503D |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Foreign Body In Patient (2687)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during an intracavitary resection procedure, the resection electrode loop had been deformed and bent.A second one was used but broke and remained stuck in the patient's uterine polyp, which the customer was unable to recover.The patient had bleeding but recuperated afterwards.No prolonged hospitalization.The procedure was delayed for ten minutes and was not completed.The patient will have a new procedure - hysterectomy.The devices were inspected before use.There were no reports of further patient or user harm associated with this event.Two reports were required for this event with the following patient identifiers: (b)(6) - 1st resection electrode loop; (b)(6) - 2nd resection electrode loop.This report is for (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the suggested event occurred due to improper handling/excessive force by the user.However, the root cause of the suggested event could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|