| Catalog Number |
405672 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tray spn whit25g3.5 l/b-d/e blue drape experienced foreign matter, contaminated.The following information was provided by the initial reporter: piece of fiber in the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up mdr for device evaluation: two photos along with the physical sample were provided to our quality team for investigation.Through visual inspection of the physical sample, a plastic fiber was observed inside the syringe, verifying the reported incident.A review of the internal manufacturing device records and raw material history files for reported lot 0001517819 was performed and no recorded quality problems or rejections related to this incident were found, all procedural and functional requirements for product release have been met.This syringe is provided by another bd site who has been notified of this incident.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the tray spn whit25g3.5 l/b-d/e blue drape experienced foreign matter, contaminated.The following information was provided by the initial reporter: piece of fiber in the syringe.
|
| |
|