• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL 3I T3 WITH DCD NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 15MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL 3I T3 WITH DCD NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 15MM; DENTAL IMPLANT Back to Search Results
Catalog Number BNST3215
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
 
Event or Problem Description
The customer reported implant failure at tooth location #26 due to infection.The implant was removed.Symptoms as a result of the event: inflammation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 WITH DCD NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 15MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17916639
Report Number0001038806-2023-01951
Device Sequence Number17810370
Product Code DZE
UDI-Device Identifier00844868029120
UDI-Public(01)00844868029120(17)260630(10)2021060518
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/29/2026
Device Catalogue NumberBNST3215
Device Lot Number2021060518
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-