• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE EXTENSION SET 15 CM (6 IN) 1.14 ML STD BORE; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD Q-SYTE EXTENSION SET 15 CM (6 IN) 1.14 ML STD BORE; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 385101
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd q-syte extension set 15 cm (6 in) 1.14 ml std bore leakage occurred there was no report of patient impact.The following information was provided by the initial reporter: unable to draw blood, air leakage.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Our quality engineer inspected the 1 sample submitted for evaluation.The reported issues of aspirate / draw (cannot / difficult) and leakage were confirmed upon inspection of the sample.During the inspection a crack was found on the extinction tubing.Bd determined that the cause of the failure was likely attributed to our supplier, however without the batch number a further review into the production records cannot be performed.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.We regret any inconveniences this incident may have caused you and your facility.H3 other text : see narrative below.
 
Event or Problem Description
No new information received unable to draw blood, air leakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Q-SYTE EXTENSION SET 15 CM (6 IN) 1.14 ML STD BORE
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17916920
Report Number9610847-2023-00281
Device Sequence Number6782577
Product Code FPA
UDI-Device Identifier00382903851010
UDI-Public(01)00382903851010
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K013621
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number385101
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer11/15/2023
Initial Report FDA Received Date10/11/2023
Supplement Report FDA Received Date11/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-