NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number 72003
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 09/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There is no evidence of device or system failure.The lack of neural response appears to be related to the positioning of the implanted components.
 
Event or Problem Description
Patient began the trial phase for the nalu peripheral nerve stimulator on (b)(6)2023.After implanting the trial leads the patient was unable to get any neural response from the system.The intention was to target the suprascapular nerve, however it was suspected that the leads were floating above the suprascapular fossa instead of hitting the desired location.A surgical revision was performed on (b)(6)2023 to remove the implanted trial leads and replace with a new set of leads targetting the desired nerve location.The trial was completed successfully and the leads were removed as planned on (b)(6)2023.Patient is preparing for permanent implant procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Common Device Name
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17917438
Report Number3015425075-2023-00241
Device Sequence Number6794774
Product Code GZF
UDI-Device Identifier00812537033631
UDI-Public0100812537033631112301241726012410UF564
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/24/2026
Device Model Number72003
Device Catalogue Number72003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2023
Initial Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-