Model Number NANONEEDLE SCOPE, 125MM
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a demo peroneus recon surgery of the nanoneedle an image scattering in the form of horizontal lines on the nanoneedle monitor was noticed.There was no harm for patient, operator or third party reported.The surgery was finished with the same device anyway.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
(no problem found) the evaluation did not identify any issues relevant to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANONEEDLE SCOPE, 125MM
Common Device Name
LAPAROSCOPE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17919508
Report Number1220246-2023-08154
Device Sequence Number2893638
Product Code GCJ
UDI-Device Identifier00888867375574
UDI-Public00888867375574
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K201134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberNANONEEDLE SCOPE, 125MM
Device Catalogue NumberAR-3210-0043
Device Lot NumberYS22010006
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2023
Supplement Date Received by Manufacturer07/03/2024
Initial Report FDA Received Date10/12/2023
Supplement Report FDA Received Date08/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-