| Model Number |
02-01894 |
| Medical Device Problem Codes |
Melted (1385); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient starting charging the monitor however the monitor charger melted.The monitor was damaged and the charging cord was very difficult to remove from the device.The patient stopped using the device and a replacement was sent.The patient stated that there was no harm or property damage.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the monitor charger melted and caused damage to the monitor.The device was returned for investigation.Engineering evaluation was able to confirm device melt.No internal device was observed during evaluation.It is most probable that the device melt was caused by an external heat source.Am&d philips is doing further investigation fro mcot usb-a/usb-c cables overheating.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.
|
| |
|