| Model Number |
02-02045 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Chest Pain (1776)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient reported that the sensor was placed at their doctor's office ontop of her implanted defibrillator.Once they got home the patient started feeling pain on their chest, legs, stomach and started experiencing a headache.The patient removed the device instantly and was suggested to seek medical attention.The patient went to the emergency room (er) due to severe chest pains.The device is expected to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported the patient experienced pain after the mcot c6 sensor was placed ontop of their defibrillator.The device was returned for investigation.Device was inspected for general physical integrity and found the blue seal on the sensor damaged from customer pry marks.Engineering evaluation could not replicate the reported event.The sensor is unlikely to have caused or contributed to the reported patient symptoms.Per the ifu, "the ecg recorder is not intended for use on patients with implanted pacemakers." a definitive root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|