MMJ SA DE CV (USD) INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number PM7100
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Image Display Error/Artifact (1304); High Readings (2459)
Health Effect - Clinical Codes Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: pm7100, monitor pm7100 tablet invos (lot#unknown); pmac71doc, docking station pmac71doc invos 7100 (serial# (b)(6)); pmpamp71, preamplifier pmpamp71 invos 7100 (serial#(b)(6)); pmac71doc, docking station pmac71doc invos 7100 (serial#(b)(6)); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmpamp71, preamplifier pmpamp71 invos 7100, serial#(b)(6) (serial#(b)(6)); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#unknown); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#unknown); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#unknown); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos k it (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot# unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown); pmsens71-a-20, adult sensor pmsens71-a invos kit (lot#unknown).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the monitor screen went suddenly gray with a system error in the middle of the ongoing surgery.The nurse anesthetist involved could not tell what error message was given.The problems occurred during hypothermic circulatory arrest and caused a massive delay in operating time.Replacing the entire system was also unsuccessful, as this monitoring system also suddenly failed.There were reports regarding not-comprehensible, incorrectly high rso2 values and a sudden change on the screen from 2 to 4channels.The nurse anesthetist noticed that there was a lot of cautery.
 
Event or Problem Description
According to the reporter, the monitor screen went suddenly gray with a system error in the middle of the ongoing surgery.The nurse anesthetist involved could not tell what error message was given.The problems occurred during hypothermic circulatory arrest and caused a massive delay in operating time.Replacing the entire system was also unsuccessful, as this monitoring system also suddenly failed.There were reports regarding incomprehensible and incorrect high rso2 values and a sudden change on the screen from 2 to 4 channels.The nurse anesthetist noticed that there was a lot of cautery in use.There was no patient injury as a result of the delay in treatment.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the cable of the unit was damaged.It was reported that the monitor displayed higher readings than expected and the display was defective.The product gave an error message but the specific value was not recorded.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the monitor screen went suddenly gray with a system error in the middle of the ongoing surgery.The nurse anesthetist involved could not tell what error message was given.The problems occurred during hypothermic circulatory arrest and caused a massive delay in operating time.Replacing the entire system was also unsuccessful, as this monitoring system also suddenly failed.There were reports regarding incomprehensible and incorrect high rso2 values and a sudden change on the screen from 2 to 4 channels.The nurse anesthetist noticed that there was a lot of cautery in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17921321
Report Number2936999-2023-00915
Device Sequence Number14198853
Product Code MUD
UDI-Device Identifier10884521552609
UDI-Public10884521552609
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K182868
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPM7100
Device Catalogue NumberPM7100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2023
Supplement Date Received by Manufacturer01/11/2024
Not provided
Initial Report FDA Received Date10/12/2023
Supplement Report FDA Received Date01/29/2024
04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H10.
Patient SexUnknown
-
-