HU-FRIEDY MFG. CO. LLC 11 STRAIGHT BERNARD SYNDESMOTOME Back to Search Results
Model Number BER11
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that during a dental procedure the user facility was using their 11 straight bernard syndesmotome when the tip detached resulting in a cut to the patient's tonsil area.The detached tip was retrieved from the patient's mouth.Medical treatment was administered (decadron administered via iv).
 
Additional Manufacturer Narrative
Hufriedy requested the device to be returned for evaluation.However, the user facility elected to keep the device until their own investigation is complete.Without the return of the device, the cause of the reported issue cannot be determined.A 3-year complaint review indicates this to be an isolated event.A follow-up mdr will be submitted should the device be returned for evaluation.
 
Additional Manufacturer Narrative
The user facility has indicated that they will not return the instrument to us for evaluation.To date, no additional information regarding the reported event has been provided.
 
Additional Manufacturer Narrative
Steris became aware of user facility medwatch report number (b)(4) through our periodic review of the maude database on (b)(6) 2023.Steris confirmed that an investigation for the reported event was previously completed and mdr # 1416605-2023-00007 for the event was submitted on (b)(6) 2023.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11 STRAIGHT BERNARD SYNDESMOTOME
Common Device Name
SYNDESMOTOME
Manufacturer (Section D)
HU-FRIEDY MFG. CO. LLC
3232 n rockwell st.
chicago IL 60618
Manufacturer (Section G)
HU-FRIEDY MFG. CO. LLC
3232 n rockwell st.
chicago IL 60618
Manufacturer Contact
daniel davy
3232 n rockwell st.
chicago, IL 60618
4403927453
MDR Report Key17923129
Report Number1416605-2023-00007
Device Sequence Number2983252
Product Code EMJ
UDI-Device Identifier10889950004767
UDI-Public10889950004767
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBER11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2023
Supplement Date Received by Manufacturer09/15/2023
09/15/2023
Initial Report FDA Received Date10/12/2023
Supplement Report FDA Received Date10/23/2023
12/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-