| Model Number |
OLH |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023 a patient underwent a posterior wall encircling ablation procedure.During repositioning/opening and closing clamp onto cardiac tissue, the blue plunger insert of the closure lever detached.The procedure was completed with the device.There was no delay in the case, nor harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The olh clamp was returned for investigation.Upon visual inspection, the blue plunger insert was missing, and the complaint was confirmed.
|
| |
|