• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72202599
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the wrong awl was employed to create access in the bone.Therefore, when the twinfix ultra anchor was tapped, it broke off.The procedure was completed with non-significant surgical delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H11: corrected information in d4 (lot number).
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with the anchor and suture strings still installed.The anchor is fractured at the proximal end of the suture window with the distal portion of the anchor held in place by the suture strings.There is biological debris on the returned items.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the raw material strength and storage requirements are specified and each lot must be accompanied by a material certificate of analysis.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17927255
Report Number1219602-2023-01972
Device Sequence Number19576147
Product Code MBI
UDI-Device Identifier03596010647474
UDI-Public03596010647474
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K093228
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202599
Device Lot Number20576
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2023
Supplement Date Received by Manufacturer10/25/2023
02/14/2024
Initial Report FDA Received Date10/12/2023
Supplement Report FDA Received Date10/25/2023
02/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-