| Catalog Number |
0502991030 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: all complaints -additional information received confirmed that the same scope could not be used to complete the procedure.Based on a review of historical complaint data, risk documentation, and the reported information regarding this event, a malfunction will be filed for blurry image.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be rough handling during sterilization and/or product transport.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|