• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 19AGFN-756
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Liver Damage/Dysfunction (1954)
Date of Event 09/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2023, a 19mm sjm regent heart valve was implant in a patient.Perioperatively, anticoagulant drug lovenox was given to the 49-year-old patient.Treatment was interrupted for 48 hours following hepatic cytolysis.Ultrasound checks were carried out for 7 days postoperatively and showed nothing and patient patient continue to have anticoagulant.The patient have any hematologic disorders or systemic illness that could contribute to a hypercoagulability.On (b)(6) 2023, it was reported that thrombosis was detected on the valve following an ultrasound check.Evidence of dysfunction of the mechanical valve with an average gradient of 67 mmhg and iao 2/4 whereas the previous ultrasounds were unremarkable on the valvular level.Transesophageal echocardiography (tee) and cine-angiography showed one wing to be immobile and another blocked in an intermediate position.The device was explanted and a non-abbott device was implanted.The patient is stable.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of thrombus on the valve was confirmed.The investigation found that there was adherent calcified thrombus on the inflow and outflow surface of the mechanical heart valve leaflets and in the pivot recesses.The leaflets were immobile.There was an adherent clot on the sewing cuff.No inflammation was present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the patient was prescribed an anticoagulant and was compliant, with no disorders or medication which could have contributed to hypercoagulability.Treatment was interrupted for 48 hours following hepatic cytolysis.Their inr at the time of the event was normal due to the progressive dose introduction of avk treatment for 4 days with coumadin.The thrombus noted on the valve could have contributed to the immobility of the valve leaflets.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17927521
Report Number2135147-2023-04491
Device Sequence Number6311863
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number19AGFN-756
Device Lot Number31210388
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/22/2023
Initial Report FDA Received Date10/12/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
-
-