| Catalog Number |
7210499 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy surgery, the disposable blade tip was blocked leaving micro fragments of metal in the joint, the pieces were removed by suction.The procedure was completed with a non-significant delay using a s+n back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The color scheme matches the print.There is scoring on the inner cannula of the outer blade.There is scoring on the outer shaft of the inner blade.Thick bio debris on both the inner and outer blades.A functional evaluation showed that manually rotating the inner blade has resistance due to the bio debris.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with an unintended use of the device.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|