| Catalog Number |
SRX-420 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported, left side capsular contracture, baker grade iii.Healthcare professional later reported, "rupture found, during explant surgery".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iii" is a physiological complication.And analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.And device rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, two openings assessed, as surgical damage.Capsular contracture: unable to observe, since it is not related to the device.Additional observations: creases are observed.Wear abrasion is observed.Nicks striated is observed.Assessed, as surgical tool scratch like.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, left side capsular contracture, baker grade iii.Healthcare professional, later reported, "rupture found, during explant surgery".Device has been explanted.
|
| |
|