• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEICA BIOSYSTEMS NUSSLOCH GMBH HISTOCORE PEGASUS; HISTOCORE PEGASUS PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEICA BIOSYSTEMS NUSSLOCH GMBH HISTOCORE PEGASUS; HISTOCORE PEGASUS PLUS Back to Search Results
Catalog Number 14048858005
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/15/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation of the incident is currently underway and a follow up will be submitted should additional information become available following the investigation.
 
Event or Problem Description
On (b)(6) 2023 leica biosystems received a complaint that the customer experienced suboptimal tissue processing on their asp6025, tissue processor.The customer informed leica biosystems, that as a result, 20 tissue samples have not been diagnosed.
 
Additional Manufacturer Narrative
The investigation revealed the following: the instrument logs were analyzed by the lbs senior quality engineer, and no instrument error was detected during the processing steps related to the damaged tissue.The instrument is working according to factory specifications.The incident was user related and caused by the fixation and usage of the incorrect assignment of the carryover value setting for created or edited protocols.If the setup of carryover is lower than the actual carryover, it may potentially cause tissue damage (mainly under-processed tissue).A customer facing letter was sent out to the customer with recommendation, to consider the carryover setting and to use the recommended accordingly the protocol for standard biopsy and small specimen as defined in the instruction for use (ifu).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOCORE PEGASUS
Common Device Name
HISTOCORE PEGASUS PLUS
Manufacturer (Section D)
LEICA BIOSYSTEMS NUSSLOCH GMBH
heidelbergerstrasse 17-19
nussloch, 69226
GM  69226
Manufacturer (Section G)
LEICA MICROSYSTEMS LTD. SHANGHAI
building 1, 258 jinzang road
shanghai,pudong, 20120 6
CH   201206
Manufacturer Contact
robert gropp
heidelbergerstr. 17-19
nussloch, 69226
GM   69226
MDR Report Key17929712
Report Number8010478-2023-00012
Device Sequence Number20433348
Product Code IEO
UDI-Device Identifier04049188216397
UDI-Public(01)04049188216397
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number14048858005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/15/2023
Supplement Date Received by Manufacturer10/20/2023
Initial Report FDA Received Date10/13/2023
Supplement Report FDA Received Date11/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-