|
The device was returned to olympus for evaluation.Upon inspection testing of the returned device, the user's request was confirmed.It was discovered, the knife wire of the distal end was blackened, fused, and separated from the ceramic head.A review of the device history record (dhr), found no deviations that could have caused or contributed to the reported issue.Based on the results of confirmation of the device and the investigation results in the past, a likely mechanism causing the broken cutting wire might be the following: 1.Due to the factor described below, spark discharge between the tissue and the cutting knife occurred during output activation.The output setting for activation was too high.The electrosurgical unit was used in the coagulation mode.The output activation time was too long.2.Spark discharge occurred, and the part of the cutting wire became hot instantly.As a result, the strength of the cutting knife decreased.Also, the cutting wire was scorched.3.During tissue incision, a load was applied to the cutting knife which has the reduced strength.This might have caused the cutting knife to break.However, the exact cause of spark discharge generation could not be identified.Therefore, the root cause of the reported event could not be determined.The instruction manual contains the following description related to this reported phenomenon.(drawing no.Gk8148, revision no.08)."when the electrosurgical unit is used in the coagulation mode, crack/detachment of the tip or deformation/break of the cutting knife or the electrode could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should cracks or detachment of the tip or deformation/break of the cutting knife be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and withdraw the endoscope from the patient with the cutting knife retreated in the insertion portion of this instrument.Do not continue using an abnormal electrosurgical knife to prevent perforation or bleeding.If the tip or cutting knife is detached be sure to collect it using grasping forceps.Aspirate fluids such as mucus that adhere to the electrode and/or the cutting knife, outer sheath and body cavity tissues.Patient injury such as punctures, hemorrhages, mucous membrane damage and thermal injury of tissue could result if output is activated when in contact with these adhering fluids.When current is discharged while the cutting knife is being separated from the mucosa under wet situation, it may break the cutting knife or crack the tip.Always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.Excessive output level and time may occur in patient injury, such as perforations, bleeding or mucous membrane damage.Application of high-voltage waveforms for extended periods increase the likelihood that it may break the cutting knife or crack the tip.When a high-voltage waveform has to be used, minimize the duration of current application.Electrosurgical unit: voltage intensities of various waveforms soft coagulation < cut / pulse cut < forced coagulation.Do not activate output continuously but activate it intermittently.Continuous activation may cause patient injury, such as bleeding, mucous membrane damage, or thermal injury of non-target tissue.Crack/detachment of the tip or deformation/break of the cutting knife or electrode may also occur.Do not activate output when the electrode and cutting knife are not in contact with the target tissue.Crack/detachment of the tip or deformation/break of the cutting knife or electrode may occur." olympus will continue to monitor the field performance of this device.
|
|
According to the reporter, the event date was september 19, 2023.The probe tip fell into the patients stomach and was retrieved using forceps.There was no need for unplanned diagnostic imaging to locate the device or confirm it was removed.The procedure was delayed by 15 minutes while the patient was under general anesthesia.There was no harm nor intervention performed due the device malfunction.There were no additional details regarding the patient or event reported.
|