|
A falsely elevated free light chains, type kappa (flc kappa) result was obtained on a patient sample on a bn ii system using a 1:100 sample dilution.The erroneous result was not reported to the physician(s).The sample was repeated two times for flc kappa, once using a 1:400 dilution and once using a 1:100 dilution, both times recovering lower.The repeat result obtained using the 1:100 dilution was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated flc kappa result.
|