| Model Number |
JAZZY PASSPORT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The charger was tested and found to be confirming.The battery pack will be evaluated by the manufacturer.A follow-up report will be forthcoming if more information becomes available.This mdr is beyond the 30-day requirement due to the fda requiring a new web trader license.There were a series of difficulties in getting this licence and numberous e-mails were sent to the fda help desk.A follow-up report will be forthcoming if more information becomes available.
|
| |
|
Event or Problem Description
|
|
Consumer alleges he smelled smoke and went to check his chair that was charging.He alleges the battery charger light was green, so he unplugged the charger cord from the chair, he allegedly heard a faint clicking sound allegedly coming from under the chair.When he looked at the chair he alleges smoke started to billow out, the battery allegedly started hissing and flames allegedly began to shoot out to about 4-6 inches in at least two directions.
|
| |
|
Event or Problem Description
|
|
Consumer alleges he smelled smoke and went to check his chair that was charging.He alleges the battery charger light was green, so he unplugged the charger cord from the chair, he allegedly heard a faint clicking sound allegedly coming from under the chair.When he looked at the chair, he alleges smoke started to billow out, the battery allegedly started hissing and flames allegedly began to shoot out to about 4-6 inches in at least two directions.
|
| |
|
Additional Manufacturer Narrative
|
|
The charger was tested and found to be confirming.The battery pack will be evaluated by the manufacturer.A follow-up report will be forthcoming if more information becomes available.This mdr is beyond the 30-day requirement due to the fda requiring a new webtrader license.There were a series of difficulties in getting this licence and numberous e-mails were sent to the fda helpdesk.A follow-up report will be forthcoming if more information becomes awailable.Device evaluaton: device sustained thermal event in battery pack.Although the exact cause could not be determined, third party testing was performed on a current battery production battery pack in which the event could not be duplicated.The passport will be discontinued at the end of the ear and is being replaced with an new model and battery pack.
|
| |
|
Search Alerts/Recalls
|