• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Sexual Dysfunction (4510)
Date of Event 06/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that the patient had presented with ejaculatory dysfunction (per manufacturer instructions for use, sexual dysfunction, including ejaculatory and erectile dysfunction are potential perioperative risks of the aquablation procedure).It was reported that the treating physician was to monitor the patient's condition.The physician determined that the ejaculatory dysfunction was related to the procedure and unlikely related to the device.No malfunction of the aquabeam robotic system was reported.
 
Additional Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Additional Manufacturer Narrative
The aquabeam robotic system is a reusable device; therefore it is currently in the possession of the user facility.The investigation of this event consisted of a review of the device history record (dhr) and labeling.A review of the device history record (dhr) ab2000-c/serial number (b)(6) was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The aquabeam robotic system instructions for use (ifu), ifu0101-00, rev.E, was reviewed and states the following: 4.3.Warnings: procedure: as with any surgical urologic procedure, potential perioperative risks of the aquablation procedure include: sexual dysfunction, including ejaculatory and erectile dysfunction.The aquabeam robotic system is a reusable device; therefore, it is still currently in possession of the user facility.The treating surgeon confirmed that the aquabeam robotic system functioned as intended, and assessed the event to be related to the procedure.The aquabeam robotic system's ifu lists sexual dysfunction, including ejaculatory and erectile dysfunction as a potential risk of the aquablation procedure.Based on the review of the dhr, and ifu the event is considered not to be device-related.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Common Device Name
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
6502327291
MDR Report Key17934955
Report Number3012977056-2023-00172
Device Sequence Number6352517
Product Code PZP
UDI-Device IdentifierB614AB2000C1
UDI-Public+B614AB2000C1/16D20220330O
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K212835
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberAB2000C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2023
Supplement Date Received by Manufacturer11/10/2023
Initial Report FDA Received Date10/13/2023
Supplement Report FDA Received Date11/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-