MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 50MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 50MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120150
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Date of Event 01/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, a plaintiff complained of bilateral hip pain, worsened in left hip, and difficulty with everyday activities; therefore, a left primary bhr was carried out on (b)(6) 2013.Then evidence of increased serum metal ion concentration was found, and the plaintiff complained of dysfunction.A left bhr revision was performed on (b)(6) 2022 to treat this event, evidence of metallosis was noticed with darkened synovial tissue, and erosion posterior to the acetabular shell.The patient was implanted a zimmer total hip system.Patient was then taken to pacu in stable condition, no other complications have been reported.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Additional Manufacturer Narrative
H3, h6.It was reported that the patient complained of bilateral hip pain, worsened in left hip, and difficulty with everyday activities; therefore, a left primary bhr was carried out.Then evidence of increased serum metal ion concentration was found, and the patient complained of dysfunction.A left bhr revision was performed to treat this event, evidence of metallosis was noticed with darkened synovial tissue, and erosion posterior to the acetabular shell.The patient was implanted a competitor total hip system.Patient was then taken to pacu in stable condition, no other complications have been reported.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the head and no other similar complaints for the cup, and this will continue to be monitored via routine trending, however it should be noted that these devices are no longer sold.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.The clinical information provided of the evidence of metallosis with darkened synovial tissue and erosion posterior to the acetabular shell may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 50MM W/ IMPTR
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17935094
Report Number3005975929-2023-00146
Device Sequence Number6455106
Product Code NXT
UDI-Device Identifier03596010502568
UDI-Public3596010502568
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/14/2014
Device Catalogue Number74120150
Device Lot Number097664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2023
Supplement Date Received by Manufacturer11/01/2023
Initial Report FDA Received Date10/13/2023
Supplement Report FDA Received Date11/03/2023
Date Device Manufactured02/15/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
74121142/RESURFACING FEMORAL HEAD 42MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexMale
-
-