| Catalog Number |
68-750 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Dhr trend summary: considering that there is not an adverse trend for this type of event no additional actions are deemed required at this time.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed, one opening assessed as unidentified (tear) opening.Additional observations: crease is observed.White particles on device inner surface are observed.No further actions are required since no issue in the manufacturing of the device is observed.
|
| |
|