| Model Number |
1005-0-25 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Skin Tears (2516)
|
| Date of Event |
09/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Left a sore when removed from babies face.Appears to have torn off layers of the babies skin.
|
| |
|
Additional Manufacturer Narrative
|
|
The baby's skin had sores and was torn when removing skin fixation.
|
| |
|
Additional Manufacturer Narrative
|
|
The baby's skin had sores and was torn when removing skin fixation.Complaint history reviewed.There have been 0 similar complaints in the previous 24 months for this part#.No product returned, no pictures of alleged defect.Complaint cannot be confirmed.Date of manufacture: (b)(6) 2023.No remaining product from lot# in inventory.Most likely root cause is skin sensitivity due to prolonged exposure to device.Rma-20011: r27: patient's skin is sensitive to device removal - age, medication, existing skin damage, individual patient sensitivity - s=7 o=4 rpn=28*.*to have an acceptable level of risk, rpn < 25 based on cpm, rrpn =7, which is an acceptable amount of risk.Rma will be corrected during fda remediation.
|
| |
|
Event or Problem Description
|
|
Left a sore when removed from babies face.Appears to have torn off layers of the babies skin.
|
| |
|