Catalog Number 367925
Medical Device Problem Code Device Ingredient or Reagent Problem (2910)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use with an unspecified amount of bd vacutainer® sodium fluoride 15mg potassium oxalate 12 mg (fx) blood collection tubes hemolysis occurs.Samples were recollected, additional patient information or impact has not been reported.The following information was provided by the initial reporter: customer reported about 40-50% of their fluoride tubes having hemolysis issue and causing specimen recollection.
 
Additional Manufacturer Narrative
D.1.Medical device brand name: bd vacutainer® sodium fluoride 15mg potassium oxalate 12 mg (fx) blood collection tubes h.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were visually inspected with no issues identified.Complaints for sample quality are under statistical control for the month of september 2023.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for hemolysis.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
 
Event or Problem Description
It was reported that during use with an unspecified amount of bd vacutainer® sodium fluoride 15mg potassium oxalate 12 mg (fx) blood collection tubes hemolysis occurs.Samples were recollected, additional patient information or impact has not been reported.The following information was provided by the initial reporter: customer reported about 40-50% of their fluoride tubes having hemolysis issue and causing specimen recollection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEE H.10
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17935943
Report Number1917413-2023-01019
Device Sequence Number12350530
Product Code JKA
UDI-Device Identifier50382903679251
UDI-Public50382903679251
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K901449
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2024
Device Catalogue Number367925
Device Lot Number3136046
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2023
Supplement Date Received by Manufacturer11/15/2023
Initial Report FDA Received Date10/13/2023
Supplement Report FDA Received Date12/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-