• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 328411
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd ultra-fine¿ insulin syringe was leaking.Verbatim: daughter of consumer reported when injecting.Insulin leaks out of the hub, where meeting the barrel of syringe.Resulting of wasting of insulin.It runs down.Stomach/site daughter of consumer reported needle hub separates at times from syringe too.
 
Additional Manufacturer Narrative
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd ultra-fine¿ insulin syringe was leaking.Verbatim: daughter of consumer reported when injecting.Insulin leaks out of the hub, where meeting the barrel of syringe.Resulting of wasting of insulin.It runs down.Stomach/site daughter of consumer reported needle hub separates at times from syringe too.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes.D9: returned to manufacturer on: 23-jan-2024.H.6.Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was not able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Event or Problem Description
It was reported that the bd ultra-fine¿ insulin syringe was leaking.The following information was provided by the initial reporter: daughter of consumer reported when injecting.Insulin leaks out of the hub, where meeting the barrel of syringe.Resulting of wasting of insulin.It runs down.Stomach/site daughter of consumer reported needle hub separates at times from syringe too.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE¿ INSULIN SYRINGE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17936490
Report Number1920898-2023-00712
Device Sequence Number20421074
Product Code FMF
UDI-Device Identifier00382908411035
UDI-Public(01)00382908411035
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K190054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2026
Device Catalogue Number328411
Device Lot Number1026137
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2023
Supplement Date Received by Manufacturer10/26/2023
01/25/2024
Initial Report FDA Received Date10/13/2023
Supplement Report FDA Received Date11/06/2023
02/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-