| Model Number |
VTICMO12.6 |
| Medical Device Problem Codes |
Misfocusing (1401); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
08/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6-device code 1494: off-label use (under 21yrs of age at date of implant).Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vticmo 12.6 implantable collamer lens of -16.50/2.0/094 (sphere/cylinder/axis) was implanted into the patient's left eye (os) on (b)(6) 2023.Refractive surprise was observed.On (b)(6) 2023 the lens was exchanged for a same model and size lens with different power and the problem was resolved.Cause reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: device evaluation: lens returned in a microcentrifuge vial in liquid.Visual observation found no visible damage to lens.Dimensional and functional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|