| Catalog Number |
EP8706H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a vascular procedure on (b)(6) 2023 and suture was used.During the procedure, the suture broke when knotting during surgery.Changed another five sutures to continue the surgery, but the same problem happened.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 10/16/2023.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: *device follow up the following additional information was received: the lot number of product w8761 should be tdbcbc.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Related reports: 2210968-2023-07770, 2210968-2023-07771, 2210968-2023-07772, 2210968-2023-07773, 2210968-2023-07774, 2210968-2023-07775.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/13/2023.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that it was received one needle-suture piece that pertain to the product code ep8706h.The product code is double-armed.During visual inspection of the returned sample, the extreme was noted to be cut probably caused by a surgical instrument.In addition, a knot was noted on the body suture.As per the conditions of the returned sample, no functional test could be performed.The other section of the suture was not returned for analysis.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-07770, 2210968-2023-07771, 2210968-2023-07772, 2210968-2023-07773, 2210968-2023-07774, 2210968-2023-07775.
|
| |
|
Search Alerts/Recalls
|
|
|