| Catalog Number |
SRL-200 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iv.Later healthcare professional reported rupture and capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: capsular contracture baker grade iv, rupture.Information contained in this report was previously submitted through psr on 01/22/2019.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iv.Later healthcare professional reported rupture and capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed 1 opening assessed as fold flaw opening.¿ capsular contracture: unable to observe as it is not related to the device.Additional observations: ¿ stress marks observed.No further actions are required as no manufacturing issues are observed.
|
| |
|