|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507510MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/16/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: rupture manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with 510cc mentor memorygel breast implants.Post-operatively, the patient experienced bilateral rupture of her prostheses, which was confirmed via ultrasound and magnetic resonance imaging.As a result, the patient underwent removal surgery on (b)(6) 2023.Follow-ups are being conducted.If more information becomes available, mentor will submit a supplemental report accordingly.This report is for the right implant.Refer to manufacturing report number 1645337-2023-12202 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
On november 07, 2023, mentor received the device for evaluation.On november 22, 2023, mentor completed a visual inspection, microscopic examination, and thickness measurement of the returned device.Device evaluation summary: visual analysis of the returned sample revealed that the breast implant was found to be ruptured.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|