|
While the vascular surgeon was deploying an icast stent in iliac artery, deployed through 7f sheath.Everything went as normal, stent deployed in iliac artery and noticed balloon detached from delivery system during removal of delivery catheter.Balloon was snared and removed through sheath.No adverse issues to patient.
|
|
The reported incident pertains to the use of an icast covered stent on a patient in the iliac artery.The details provided indicate that the surgeon noticed that after normal deployment of the stent in the iliac artery, the balloon detached from the delivery system during deflation.This led to the need to snare the balloon.The device in question, an icast covered stent 9mmx59mmx80cm manufactured in february 2022, and the introducer sheath, a cook 7fr balkin, were returned and evaluated to determine the cause of the complaint.The balloon of the catheter had been separated from the catheter shaft at the location of the proximal balloon weld confirming the claim of a component separation.The weld had been elongated prior to the balloon separation and is indicative of tensile loading of the shaft.The balloon was in a curved configuration and appeared to be in good condition.To determine if the balloon had ruptured during the procedure which may have prevented the balloon from deflating, the weld area of the balloon was placed inside a toughy borst adapter.This adapter was then attached to a 20cc insufflator, and the balloon was pressurized.There was no evident leak in the balloon.A vacuum was then pulled on the balloon and the balloon was able to be fully deflated.There was no evidence of a leak during the process of inflation of the balloon fluid was easily seen entering the balloon via the two skive holes under the balloon indicating that both lumens were patent.One is on the proximal end of the balloon and one on the distal end of the balloon.The two skive holes correspond to the inflation skives within the inflation manifold where the inflation device is attached to inflate the balloon and deploy the stent.When evaluated, the skive holes within the inflation manifold were also both patent.The patency of these skive holes is critical not only to the deployment of the stent but also the deflation of the balloon after stent deployment.The investigation has shown the catheter was patent and would have deflated fully under normal circumstances.There was no damage to the catheter shaft noted during the investigation other than the proximal weld separation.The introducer sheath used in the case was also returned.The sheath was a cook 7fr balkin introducer sheath.The tip of the sheath was deformed.There were also three locations along the sheath where it appears as if the sheath outer cover collapsed upon itself.This is seen at the distal end of the sheath as well as the proximal end of the sheath.It is not known if the collapse of the sheath was related to an attempted withdrawal of the catheter or if the sheath was damaged prior.A thorough review of the device history records was performed, which found no deviations, nonconformances, or anomalies during manufacture.All test samples met performance testing requirements, including tensile testing of the catheter shaft.There is no evidence that a manufacturing defect caused or contributed to the reported failure.Complaint trending found that the actual occurrence level did not exceed the anticipated occurrence level and the reported defect is sufficiently covered by the risk documentation.There is evidence of a large tensile load having been applied to the catheter shaft where the balloon broke off, as evidenced by the necked down catheter shaft near the breaking point.This tensile load is considered to have been the cause of the detachment.It is not known when or why this tensile load was applied to the delivery system based on the device evaluation and user feedback.However, according to the device risk assessment, the most significant factor contributing to component separation is fluid remaining in the balloon during removal due to insufficient deflation of the balloon prior to withdrawal.Force applied to the delivery system in an attempt to remove the device which has not been fully deflated can cause the catheter shaft to break, causing the balloon and/or hub to separate from the rest of the delivery system.The component separation and inability to deflate the balloon have not been confirmed to have been directly related to a deficiency in the product and there is no indication of any nonconformance present prior to the attempted deflation.There is no deficiency identified in the design labeling, and risk documentation nor has a manufacturing problem been identified.The root cause of the balloon detachment is operational context of the procedure combined with user error.
|