| Model Number |
H700488 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrioventricular node ablation procedure, output issues requiring troubleshooting measures resulted in a procedural delay.The generator and catheter were replaced and the procedure was completed with no adverse consequences to the patient.The ablation catheter was inserted into the patient but an error message on the ampere generator read "generator malfunction" and then read "catheter not connected".The cable was replaced twice, and the catheter was exchanged.The ampere was rebooted, but the error message remained.After troubleshooting, it was determined that the issue was with the ampere.The procedure was completed using a non-abbott (biosense webster) generator and catheter.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One ampere¿ rf ablation generator mn h700488 and sn (b)(6) was received for evaluation.Based on the information provided to abbott and the investigation performed, the reported event was unable to be confirmed.The reported product experience observed in the field could not be reproduced in the lab.A review of the engineering logs revealed an ¿ablation control thermalboard power on self test failed¿ errors that corresponded to the reported product experience.The returned product functioned properly during the evaluation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|
| |
|