LIFE SPINE PROLIFT MICRO; INTERVERTEBRAL BODY FUSION DEVICE Back to Search Results
Model Number 58-0832-0813M
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code Unspecified Musculoskeletal problem (4535)
Date of Event 09/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The prolift micro expandable spacer system was unable to be returned to life spine, therefore, it is not possible to determine what was the complete failure mode resulting in the reported event.However, based on the description of the event it is hypothesized that the root cause of this failure was likely loosened set screw in the pedicle screw system (pedicle screw system not a lifespine product) leading to compromised spinal stability and therefore resulting in the flipping of the prolift micro expandable.
 
Event or Problem Description
It was reported that: " [we] were notified by the surgeon and distributor that the prolift micro flipped over.They think it was due to the set screw from another pedicle screw system that unattached.Therefore, creating an unstable construct.Revision surgery to be scheduled the week of sept 18th".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLIFT MICRO
Common Device Name
INTERVERTEBRAL BODY FUSION DEVICE
Manufacturer (Section D)
LIFE SPINE
13951 quality drive
huntley IL 60142
Manufacturer Contact
angela batker
13951 quality drive
huntley 60142
8478846117
MDR Report Key17942066
Report Number3004499989-2023-00020
Device Sequence Number2982952
Product Code MAX
UDI-Device Identifier00190837141684
UDI-Public(01)00190837141684(17)260608(10)JZ61
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
D453301K2125
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/08/2026
Device Model Number58-0832-0813M
Device Lot NumberJZ61
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2023
Initial Report FDA Received Date10/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-