• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PRC AGMT BLOCK DIST SZ D 5MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PRC AGMT BLOCK DIST SZ D 5MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately 5 years post implantation, the patient was revised due to pain and instability.No obvious cause for instability was identified during the revision procedure.Attempts have been made and no additional information is available.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not returned by the facility.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted g2: australia.Additional associated products & mdrs.------------------------------------------------------ 00598800300 size 3 precoat cemented stemmed a/p wedged tibial lot# 63885791 mdr: 0001822565-2023-02872.00598802012 12mm diameter 100mm length offset stem ext combined length 145mm lot# 63981070 mdr: 0001822565-2023-02873.00598802013 13mm diameter 100mm length offset stem ext combined length 145mm lot# 63941617 mdr: 0001822565-2023-02874.00599401491 left size d cemented option femoral lot# 63850332 mdr: 0001822565-2023-02876.00599403012 2mm height size c,d with locking screw yellow articular surface lot# 63713247 mdr: 0001822565-2023-02877.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned provided pictures identified explanted implants which were covered in blood.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: hinged left total knee arthroplasty with no hardware failure or loosening.No fracture seen.Complaint cannot be confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRC AGMT BLOCK DIST SZ D 5MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17943959
Report Number0001822565-2023-02875
Device Sequence Number7822959
Product Code JWH
UDI-Device Identifier00889024563865
UDI-Public(01)00889024563865(17)270531(10)63683407
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K946150
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00599003410
Device Lot Number63683407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2023
Supplement Date Received by Manufacturer01/29/2024
Initial Report FDA Received Date10/16/2023
Supplement Report FDA Received Date02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight66 KG
-
-