| Model Number |
401771 |
| Medical Device Problem Code |
Material Too Rigid or Stiff (1544)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a procedure for a patient with st-segment elevation myocardial infarction with complete block, a pericardial effusion occurred which required a pericardiocentesis and surgery to stabilize the patient.The pacing catheter was positioned as temporary and a stent was placed on the circumflex coronary artery, when a drop in blood pressure was observed.An echocardiogram was performed and confirmed a pericardial effusion.A pericardiocentesis was done, but bleeding continued because the temporary device moved and created a small hole in the wall.The patient was sent to surgery in parma.The presence of the hole in the ventricular wall was confirmed and the patient stabilized.The physician believes the tip of the catheter was too stiff which may have caused the perforation.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|